Search    
EN NL FR

  

Automation has a complete range of single phase medical transformers of the type MED-1 compliant to the applicable medical standards :

 

IEC 61558-2-15 / EN 61558-2-15 / IEC 61558-1

 

• Available models : from 550VA to 10 kVA

 

• Input voltage : 230V

 

• Output voltage (no-load) : 115V - 0 - 115V

 

• Frequency : 50/60 Hz

 

• Short circuit voltage : Max. 3% Unom prim

 

• No load current : Max. 3% Inom prim

 

• Start up current : Max. 8 x Ipeak prim

 

• Leak current between secundary - earth : Max. 50 µA

 

• Leak current between case - earth : Max. 500 µA

 

• Isolating resistance : > 200 MΩ

 

• Primary and secundary protection : possible on request

 

• Standard electrostatic screen between primary and secundary winding

 

• Temperature detection : standard equiped with PTC. Option: PT100

 

• Protection case IP21 : possible on request 

 

• Available in economic and extra economic version.


• Triple phase transformers: on request

 

 

 

 

 

 

  

 

  

Technical datasheet

Not available in English. Check Dutch or French pages or contact our commercial department.