Automation has a complete range of single phase medical transformers of the type MED-1 compliant to the applicable medical standards :
IEC 61558-2-15 / EN 61558-2-15 / IEC 61558-1
|
• Available models : from 550VA to 10 kVA
• Input voltage : 230V
• Output voltage (no-load) : 115V - 0 - 115V
• Frequency : 50/60 Hz
• Short circuit voltage : Max. 3% Unom prim
• No load current : Max. 3% Inom prim
• Start up current : Max. 8 x Ipeak prim
• Leak current between secundary - earth : Max. 50 µA
• Leak current between case - earth : Max. 500 µA
• Isolating resistance : > 200 MΩ
• Primary and secundary protection : possible on request
• Standard electrostatic screen between primary and secundary winding
• Temperature detection : standard equiped with PTC. Option: PT100
• Protection case IP21 : possible on request
• Available in economic and extra economic version.
• Triple phase transformers: on request
|